Published 2012 | Version v1
Journal article

Metal micro-detectors: development of 'transparent' position sensitive detector for beam diagnostics

  • 1. Kiev Institute for Nuclear Research, Kyiv (Ukraine)

Description

Metal Micro-strip Detector (MMD) represents a novel position sensitive detector for wide range of applications. The main advantages of MMD are low thickness, high radiation resistance and high spatial resolution. MMD production technology includes some stages: micro-strip layout made by photo-lithography on silicon wafer, plasma-chemistry etching of the silicon wafer in the operating window, micro-cabling connection to the readout electronics and DAQ. Commercially available read-out systems (VASCM3 microchip preamplifier, Time Pix readout chip, Gotthard, X-DAS) have been studied for use with MMD. Characterization studies of the MMD are presented in details.

Additional details

Publishing Information

Journal Title
Voprosy Atomnoj Nauki i Tekhniki
Journal Issue
no.6-82/18
Journal Page Range
p. 196-198
ISSN
1562-6016