An investigation has been made of the count-rate response of several CdTe detectors, to dose-rates in the range 0.1 Rad/h to 80 Rad/h. The relationship was found to be non-linear and it is suggested that this is due to a reduction in the depletion layer thickness resulting from the trapping of charge carriers. Over a limited range of dose-rates the relationship is of the form C = KDsup(n), with a mean value of n of 0.77 +- 0.03. This value is consistent with the limits of n of 0.67 and 1, predicted by a model proposed for the reduction in depletion layer thickness. Radiation induced polarization was also observed but is thought not to be the cause of the non-linear response. (Auth.)
Additional details
Publishing Information
- Journal Title
- Nucl. Instrum. Methods
- Journal Volume
- 164
- Journal Issue
- 3
- Series
- Nucl. Instrum. Methods.
- Journal Page Range
- 469-475
- ISSN
- 0029-554X
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 11503047
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CDTE SEMICONDUCTOR DETECTORS; CHARGE CARRIERS; COUNTING RATES; DOSE RATES; DOSE-RESPONSE RELATIONSHIPS; GAMMA DETECTION; NONLINEAR PROBLEMS; POLARIZATION; TRAPPING
- Descriptors DEC
- MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS