Electron pencil-beam redefinition algorithm dose calculations in the presence of heterogeneities
- 1. Department of Radiation Physics, University of Texas M.D. Anderson Cancer Center, 1515 Holcombe Boulevard, Houston, Texas 77030 (United States)
Description
The electron pencil-beam redefinition algorithm (PBRA) is currently being refined and evaluated for clinical use. The purpose of this work was to evaluate the accuracy of PBRA-calculated dose in the presence of heterogeneities and to benchmark PBRA dose accuracy for future improvements to the algorithm. The PBRA was evaluated using a measured electron beam dose algorithm verification data set developed at The University of Texas M. D. Anderson Cancer Center. The data set consists of measurements made using 9 and 20 MeV beams in a water phantom with air gaps, internal air and bone heterogeneities, and irregular surfaces. Refinements to the PBRA have enhanced the speed of the dose calculations by a factor of approximately 7 compared to speeds previously reported in published data; a 20 MeV, 15x15 cm2 field electron-beam dose distribution took approximately 10 minutes to calculate. The PBRA showed better than 4% accuracy in most experiments. However, experiments involving the low-energy (9 MeV) electron beam and irregular surfaces showed dose differences as great as 22%, in albeit a small fractional region. The geometries used in this study, particularly those in the irregular surface experiments, were extreme in the sense that they are not seen clinically. A more appropriate clinical evaluation in the future will involve comparisons to Monte Carlo generated patient dose distributions using actual computed tomography scan data. The present data also serve as a benchmark against which future enhancements to the PBRA can be evaluated
Additional details
Identifiers
- DOI
- 10.1118/1.1406521;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 28
- Journal Issue
- 10
- Journal Page Range
- p. 2096-2104
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 35001756
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COMPUTERIZED TOMOGRAPHY; DOSIMETRY; ELECTRON BEAMS; ELECTRON DOSIMETRY; MONTE CARLO METHOD; POSITRON DETECTION
- Descriptors DEC
- BEAMS; CALCULATION METHODS; CHARGED PARTICLE DETECTION; DETECTION; DIAGNOSTIC TECHNIQUES; DOSIMETRY; LEPTON BEAMS; PARTICLE BEAMS; RADIATION DETECTION; TOMOGRAPHY
Optional Information
- Notes
- (c) 2001 American Association of Physicists in Medicine.