Published October 1977 | Version v1
Journal article

Standards in medical neutron dosimetry

Creators

  • 1. Radiobiological Inst. TNO, Rijswijk, Netherlands

Description

Clinical and radiobiological experience has shown that differences in absorbed dose of less than 10 percent can be recognized for tumor eradication probability as well as for normal tissue damage. These observations determine the degree of precision and overall accuracy required in medical neutron dosimetry. Standard neutron fields are available only for restricted conditions; the feasibility of a number of detectors as standard dosimeters is considered. Uncertainties in basic physical parameters are discussed, and recommendations for future research in neutron dosimetry for biology and medicine are included. 4 figures, 6 tables

Additional details

Publishing Information

Journal Title
Natl. Bur. Stand. (U.S.), Spec. Publ.
Journal Issue
no.493
Series
Natl. Bur. Stand. (U.S.), Spec. Publ.

Conference

Title
NBS international specialists symposium on neutron standards and applications.
Dates
28 Mar 1977.
Place
Gaithersburg, MD, USA.

Optional Information

Notes
See CONF-770321--.