Published November 26, 2007 | Version v1
Journal article

Accuracy of Measured Activity by the Dual Head Gamma Camera

  • 1. Department of Nuclear Medicine, Medical University Lodz, ul. Czechoslowacka 8/10, 92-216, Lodz (Poland)
  • 2. Department of Nuclear Physics and Radiation Safety, University of Lodz, ul. Pomorska 149/153, 90-236 Lodz (Poland)

Description

The subject of research was investigation of the accuracy in determining the 99mTc activity injected in a container simulating a single organ. Data acquisition as made with different width and range of energy windows described in the article. The activities of 99mTc were determined with an error ranging from 7% to 27%. The most probable source of errors are scattered photons

Additional details

Identifiers

Publishing Information

Journal Title
AIP Conference Proceedings
Journal Volume
958
Journal Issue
1
Journal Page Range
p. 280-281
ISSN
0094-243X
CODEN
APCPCS

Conference

Title
4. International summer school on nuclear physics methods and accelerators in biology and medicine
Dates
8-19 Jul 2007
Place
Prague (Czech Republic)

Optional Information

Notes
(c) 2007 American Institute of Physics