This work performs the performance evaluation of 55 Brazilian nuclear medicine services in activity measurement of radiopharmaceutical containing 99Tcm. Proficiency tests based on the acceptance criteria of the Regulation ISO/IEC Guide 43-1 and on the accuracy of the brazilian Regulation CNEN-NN-3.05 were applied in 63 results of activity measurements in radionuclide calibrators used by those services. The performance services has shown that the criteria of the ISO/IEC 43-1, nevertheless to be more restrictive, presents results very consistent with the accuracy criteria requested by the Brazilian regulation