Pulse counting technique and mass filter
Creators
- 1. Electrotechnical Lab., Tanashi, Tokyo (Japan). Tanashi Branch
Description
Technical and practical problems are discussed in combining the ion-pulse counting method with a quadrupole mass filter. To eliminate the interference of high frequency voltage applied to the quadrupole rods, an appropriate combination of time constant values was chosen for the CR-RC circuit in the linear amplifier of an usual counting system. Tests with a selected combination have shown that extremely low ion currents less than 1-0.1 counts per sec(cps) are possible to be measured with a background counts around 0.02-0.05 cps. With this measuring condition, residual gas analysis of 10-7 Torr can be made by using only 10-11 amp. of electron bombarding currents. Some other interesting results which were obtained with the low bombarding currents are also described.
Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of the Mass Spectrometry Society of Japan
- Journal Volume
- 20
- Journal Issue
- 3
- Series
- Shitsuryo Bunseki.
- Journal Page Range
- 197-204
- ISSN
- 1884-3271
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 4059406
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COUNTING RATES; FILTERS; IONIZATION POTENTIAL; MASS SPECTROSCOPY; PREAMPLIFIERS; PULSE AMPLIFIERS; PULSE ANALYZERS; PULSE TECHNIQUES; PULSES; SIGNALS
- Descriptors DEC
- AMPLIFIERS; ELECTRONIC EQUIPMENT; SPECTROSCOPY