Published May 1987 | Version v1
Journal article

Dosimetric substrantiation of the joining of fields of complex shape in total irradiation of the lymphatic collectors

  • 1. Tsentral'nyj Nauchno-Issledovatel'skij Rentgeno-Radiologicheskij Inst., Leningrad (USSR)

Description

The paper is concerned with the results of experimental dosimetric studies on an anthropomorphous phantom using thermoluminiscent dosimetry at the site of the site of the joining of large fields of complex shape simulating total irradiation of the lymphatic collectors. The studies demonstrated the appropriateness of the joining at a distance of 1 cm between the fields. In such a variant adequate irradiation of the lymphatic system was achieved whereas an increase in a distance between the fields resulted in a risk of recurrence in underirradiated areas without a significant decrease in radiation dose exposure to healthy organs and tissues

Additional details

Additional titles

Original title (Russian)
Дозиметрическое обоснование стыковки крупных полей сложной конфигурации при тотальном облучении лимфатических коллекторов

Publishing Information

Journal Title
Med. Radiol.
Journal Volume
32
Journal Issue
5
Series
Med. Radiol.
Journal Page Range
24-27
ISSN
0025-8334
CODEN
MERAA