Published June 2012
| Version v1
Journal article
A research of the X-rays energy response of ionization chamber dosimeter used in radiation therapy
Description
The X-rays energy response of three types of ionization chamber dosimeter was researched used by the radiation beam of the primary exposure standard for medium energy X-rays (60∼250 kV). Ten instruments of every type dosimeter were calibrated, the average value and the standard deviation of energy response were obtained, and we also analyzed and evaluated the results. The energy response of two types of ionization chamber is below 4%, it is suitable for application in radiotherapy according to the verification regulation of ionization chamber dosimeters used in radiotherapy. (authors)
Additional details
Publishing Information
- Journal Title
- Nuclear Electronics and Detection Technology
- Journal Volume
- 32
- Journal Issue
- 6
- Journal Page Range
- p. 675-677
- ISSN
- 0258-0934
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 49053827
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- DATA; DOSEMETERS; IONIZATION; IONIZATION CHAMBERS; RADIOTHERAPY; VERIFICATION; X RADIATION
- Descriptors DEC
- ELECTROMAGNETIC RADIATION; INFORMATION; IONIZING RADIATIONS; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; RADIATION DETECTORS; RADIATIONS; RADIOLOGY; THERAPY
Optional Information
- Notes
- 2 figs., 3 tabs., 6 refs.