Published June 1985
| Version v1
Report
Open
Bolometers as spectrometers for X-ray astronomy
Description
We have investigated the possiblity to use composite diamond-germanium bolometers, originally developed for I.R. astronomy, for the individual detection of X-ray photons. The advantage of a composite bolometer is that it separates the absorption and detection functions. Diamond, as absorber, is especially interesting because of its high-thermal diffusivity and low heat capacity. Associated with a 103 cm2 telescope such a device, whose performance is essentially signal limited, is able to detect, an emission line of 10-7 ph.cm-2s-1 in 105 seconds in the spectrum of a faint X-ray source
Availability note (English)
MF available from INIS under the Report Number.Files
17040980.pdf
Files
(153.6 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:84f45fb45e6912cd37096ecac00dd880
|
153.6 kB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 4 p.
- Report number
- CEA-CONF--8045
Conference
- Title
- Workshop on X-ray spectroscopy.
- Dates
- 24-26 Jun 1985.
- Place
- Lyngby (Denmark).
INIS
- Country of Publication
- France
- Country of Input or Organization
- France
- INIS RN
- 17040980
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BOLOMETERS; COSMIC X-RAY SOURCES; DIAMONDS; GERMANIUM; SPECIFIC HEAT; TELESCOPES; THERMAL DIFFUSIVITY; X-RAY SPECTROMETERS
- Descriptors DEC
- CARBON; COSMIC RAY SOURCES; ELEMENTS; MEASURING INSTRUMENTS; METALS; MINERALS; NONMETALS; PHYSICAL PROPERTIES; SPECTROMETERS; THERMODYNAMIC PROPERTIES