Published 1983
| Version v1
Report
Open
Study of the characteristics of multiwire proportional chambers using cathode data read-out
Creators
- 1. Joint Inst. for Nuclear Research, Dubna (USSR)
- 2. Moskovskij Inzhenerno-Fizicheskij Inst. (USSR)
- 3. Gosudarstvennyj Komitet po Ispol'zovaniyu Atomnoj Ehnergii SSSR, Serpukhov. Inst. Fiziki Vysokikh Ehnergij
Description
The characteristics of multiwire proportional chambers using cathode data read-out have been studied. Spatial resolution of 37 μm has been obtained at the width of cathode strips of 6 mm and a 4.8 mm gap between anode and cathode. The appearance of a self quenching streamer discharge in the proportional chamber results in worsening of the spatial resolution. The chamber time resolution is 16 ns (FWHM)
Availability note (English)
MF available from INIS under the Report Number.Files
16014994.pdf
Files
(167.7 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:c3351ca6458471bf0ac6358c1f3ab196
|
167.7 kB | Preview Download |
System files
(19.4 kB)
| Name | Size | Download all |
|---|
Additional details
Additional titles
- Original title (Russian)
- Исследование характеристик многопроволочных пропорциональных камер с катодным сьемом информации
Publishing Information
- Imprint Pagination
- 8 p.
- Report number
- IFVE-OEF--83-182
INIS
- Country of Publication
- USSR
- Country of Input or Organization
- USSR
- INIS RN
- 16014994
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BETA DETECTION; EFFICIENCY; MULTIWIRE PROPORTIONAL CHAMBER; SPATIAL RESOLUTION; SPECIFICATIONS; TIME RESOLUTION
- Descriptors DEC
- CHARGED PARTICLE DETECTION; DETECTION; MEASURING INSTRUMENTS; PROPORTIONAL COUNTERS; RADIATION DETECTION; RADIATION DETECTORS; RESOLUTION; TIMING PROPERTIES
Optional Information
- Notes
- 8 refs.; 7 figs.; submitted to the journal Instrum. Exp. Tech.