Description

A new type of mini multi-element tissue-equivalent proportional counter (TEPC) based on a gas electron multiplier (GEM) has been designed and constructed. This counter is in particular suitable to be constructed with a small sensitive volume so that it can be used for microdosimetry in intense pulsed radiation fields to measure the microdosimetric spectrum in the beam of, for instance, a clinical linear accelerator. The concept lends itself also for a mini multi-element version of the counter to be used for applications in which a high sensitivity is required. In this paper, we present the first microdosimetric measurements of this novel counter exposed to a 14 MeV monoenergetic neutron beam and a californium (252Cf) source for a counter cavity diameter of 1.8 mm simulating 1.0 μm tissue site size. The measured spectra showed an excellent agreement with spectra from the literature. The specific advantages of the TEPC-GEM are discussed. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/nch144

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
110
Journal Issue
1-4
Journal Page Range
p. 839-843
ISSN
0144-8420

Conference

Title
9. Symposium on Neutron Dosimetry - Advances in nuclear particle dosimetry for radiation protection and medicine
Dates
28 Sep - 3 Oct 2003
Place
Delft (Netherlands)

Optional Information

Notes
9 refs.