SU-E-T-104: Development of 3 Dimensional Dosimetry System for Gamma Knife
Description
Purpose: The aim of this study was to develop a new 3 dimensional dosimetry system to verify the dosimetric accuracy of Leksell Gamma Knife-Perfexion™ (LGK) (Elekta, Norcross, GA). Methods: We designed and manufactured a lightweight dosimetry instrument to be equipped with the head frame to LGK. It consists of a head phantom, a scintillator, a CCD camera and a step motor. The 10×10 cm2 sheet of Gd2O3;Tb phosphor or Gafchromic EBT3 film was located at the center of the 16 cm diameter hemispherical PMMA, the head phantom. The additional backscatter compensating material of 1 cm thick PMMA plate was placed downstream of the phosphor sheet. The backscatter plate was transparent for scintillation lights to reach the CCD camera with 1200×1200 pixels by 5.2 um pitch. With This equipment, 300 images with 0.2 mm of slice gap were acquired under three collimator setups (4mm, 8mm and 16mm), respectively. The 2D projected doses from 3D distributions were compared with the exposured film dose. Results: As all doses normalized by the maximum dose value in 16 mm setup, the relative differences between the equipment dose and film dose were 0.2% for 4mm collimator and 0.5% for 8mm. The acquisition of 300 images by the equipment took less than 3 minutes. Conclusion: The new equipment was verified to be a good substitute to radiochromic film, with which required more time and resources. Especially, the new methods was considered to provide much convenient and faster solution in the 3D dose acquisition for LGK
Additional details
Identifiers
- DOI
- 10.1118/1.4888434;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 41
- Journal Issue
- 6
- Journal Page Range
- p. 246
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 46046176
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S60: APPLIED LIFE SCIENCES;
- Descriptors DEI
- ACCURACY; CAMERAS; CHARGE-COUPLED DEVICES; COLLIMATORS; DOSIMETRY; ELECTRIC MOTORS; HEAD; IMAGES; PHANTOMS; PMMA; RADIATION DOSES
- Descriptors DEC
- BODY; DOSES; ELECTRICAL EQUIPMENT; ENGINES; EQUIPMENT; ESTERS; MOCKUP; MOTORS; ORGANIC COMPOUNDS; ORGANIC POLYMERS; POLYACRYLATES; POLYMERS; POLYVINYLS; SEMICONDUCTOR DEVICES; STRUCTURAL MODELS
Optional Information
- Notes
- (c) 2014 American Association of Physicists in Medicine