Published April 2018
| Version v1
Journal article
A method for self-attenuation and sample-height correction for counting efficiency of HPGe using Marinelli beaker geometry
- 1. K.N. Toosi University of Technology, Tehran (Iran, Islamic Republic of). Department of Physics
- 2. Amirkabir University, Tehran (Iran, Islamic Republic of). Nuclear Engineering Department
Description
A procedure for self-attenuation and sample height correction in HPGe gamma spectrometry efficiency has been presented. An MCNP model of an HPGe detector was used to calculate the full energy peak efficiency (FEPE) for a group of different samples with different heights in Marinelli beaker geometry. A proper function has been fitted to the simulation results to obtain the correction function. The function has been used to calculate the FEPE of a spiked soil sample in different sample heights by considering the experimentally known FEPE of another standard solution source. A good agreement between the experiments and calculations have been shown. (author)
Additional details
Publishing Information
- Journal Title
- Journal of Radioanalytical and Nuclear Chemistry
- Journal Volume
- 316
- Journal Issue
- 1
- Journal Page Range
- p. 129-137
- ISSN
- 0236-5731
- CODEN
- JRNCDM
INIS
- Country of Publication
- Hungary
- Country of Input or Organization
- Hungary
- INIS RN
- 49052898
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ANALYTICAL SOLUTION; CALIBRATION; CORRECTIONS; COUNTING RATES; CYLINDERS; EFFICIENCY; GAMMA SPECTROSCOPY; GEOMETRY; HIGH-PURITY GE DETECTORS
- Descriptors DEC
- GE SEMICONDUCTOR DETECTORS; MATHEMATICAL SOLUTIONS; MATHEMATICS; MEASURING INSTRUMENTS; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; SPECTROSCOPY
Optional Information
- Notes
- 24 refs.