Problems in signal-to-noise ratio for attenuation correction in high resolution PET
Creators
- 1. Div. of Nuclear Medicine and Biophysics, Dept. of Radiological Sciences, UCLA School of Medicine, Los Angeles, CA 90024
Description
In high resolution PET, the number of events required for a high signal-to-noise ratio in the measured attenuation correction is an order of magnitude higher (50-100 million) than that required for the emission image. This is due to the large number of elements in the attenuation correction sinogram (e.g. 50,000) and the large attenuation factors through the thicker sections of the body (e.g. 50). Another major contribution to noise in high resolution PET is contamination of true coincidences by accidentals and their subtraction from the prompt coincidences. In this work three methods which reduce noise in the measured attenuation are evaluated. The methods are: (1) optimization of energy threshold, (2) processing of random sinogram before subtraction and (3) spatial averaging of the transmission sinogram before performing attenuation correction
Additional details
Publishing Information
- Journal Title
- IEEE Trans. Nucl. Sci.
- Journal Volume
- NS-34
- Journal Issue
- 1
- Series
- IEEE Trans. Nucl. Sci.
- Journal Page Range
- 288-293
- ISSN
- 0018-9499
- CODEN
- IETNA
Conference
- Title
- Nuclear science and nuclear power systems symposium.
- Dates
- 29-31 Oct 1986.
- Place
- Washington, DC (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 18060730
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ATTENUATION; CORRECTIONS; DATA PROCESSING; DIAGRAMS; IMAGES; POSITRON COMPUTED TOMOGRAPHY; RESOLUTION; SIGNAL-TO-NOISE RATIO; THICKNESS; THRESHOLD ENERGY
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; DIMENSIONS; EMISSION COMPUTED TOMOGRAPHY; ENERGY; INFORMATION; TOMOGRAPHY
Optional Information
- Secondary number(s)
- CONF-861007--.