Characterization of neutron fields around high energy X-ray radiotherapy machines

  • 1. Czech Metrological Institute, Prague (Czech Republic)

Description

The photoneutron spectra around a SATURNE 43 linear accelerator were characterized using a Bonner sphere spectrometer with a small 3He proportional counter. The summary neutron ambient dose equivalent values range roughly from 100 to 150 μSv/Gy; the percentages for the groups of 10-9 to 10-3 MeV, 10-3 to 1 MeV, and 1-20 MeV at 50 cm, 100 cm, 150 cm, and maze entry are tabulated. It is concluded that the neutron ambient dose equivalent to the whole body as derived from the observed neutron spectral fluence amounts approximately to 10-4 of the X-ray dose. (P.A.)

Files

30031532.pdf

Files (96.3 kB)

Name Size Download all
md5:0f318ca81ed78ec8729a675d6fdbff4b
96.3 kB Preview Download
Part of:
Proceedings of the IRPA regional symposium on radiation protection in neighbouring countries of Central Europe

Additional details

Publishing Information

Imprint Title
Proceedings of the IRPA regional symposium on radiation protection in neighbouring countries of Central Europe
Imprint Pagination
671 p.
Journal Page Range
p. 254-256
Report number
INIS-CZ--0021

Conference

Title
IRPA regional symposium on radiation protection in neighbouring countries of Central Europe
Dates
8-12 Sep 1997
Place
Prague (Czech Republic)

INIS

Country of Publication
Czech Republic
Country of Input or Organization
Czech Republic
INIS RN
30031532
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BONNER SPHERE SPECTROMETERS; DOSE EQUIVALENTS; LINEAR ACCELERATORS; NEUTRON SPECTRA; NEUTRONS; RADIOTHERAPY
Descriptors DEC
ACCELERATORS; BARYONS; ELEMENTARY PARTICLES; FERMIONS; HADRONS; MEASURING INSTRUMENTS; MEDICINE; NEUTRON SPECTROMETERS; NUCLEONS; SPECTRA; SPECTROMETERS; THERAPY

Optional Information

Notes
2 tabs., 2 figs., 5 refs.