A new digital image enhancement for the verification of field placement
- 1. Centre de Lutte Contre le Cancer Claudius-Regaud, 31 - Toulouse (France)
Description
To increase geometric treatment accuracy in radiation therapy, we used a novel digitized method and original image processing. The radiographic films that are conventionally used for verifying each beam during the treatment were digitized by a kodak digital system and then an original image enhancement was applied. For the evaluation of our technique, a clinical trial with two tests was used. The trial involved four readers doing 80 reading. The enhancement films were judged to be of higher quality than the non enhancement films(p=0.001) and were read more accurately(p ≤ 0.001). This automatic enhancement of digitized captured portal images can be easily integrated into the busy routine of a radiotherapy department. (Author), 27 refs., 3 figs., 2 tabs., 2 photos
Additional details
Additional titles
- Original title (French)
- Apport d'une nouvelle technique de numerisation au controle des champs d'irradiation
Publishing Information
- Journal Title
- Bulletin du Cancer. Radiotherapie
- Journal Volume
- 82
- Journal Issue
- 4
- Journal Page Range
- p. 379-387.
- ISSN
- 0924-4212
- CODEN
- BCRAEE
INIS
- Country of Publication
- France
- Country of Input or Organization
- France
- INIS RN
- 27053777
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- DIGITAL SYSTEMS; EXTERNAL IRRADIATION; IMAGES; NEOPLASMS; POSITIONING; QUALITY CONTROL; RADIOTHERAPY
- Descriptors DEC
- CONTROL; DISEASES; IRRADIATION; MEDICINE; THERAPY