Evaluation of the image registration program for portal images using CR and DRR images in radiation therapy

  • 1. Showa University, Graduate School of Health Sciences, Tokyo (Japan)
  • 2. Kyorin University, Faculty of Health Sciences, Tokyo (Japan)
  • 3. Showa University, Northern Yokohama Hospital, Department of Radiology, Yokohama, Kanagawa (Japan)
  • 4. Tsukuba International University, Faculty of Medical and Health Sciences, Tsuchiura, Ibaraki (Japan)
  • 5. Showa University, Fujigaoka Hospital, Yokohama, Kanagawa (Japan)
  • 6. Tokyo Metropolitan University, Tokyo (Japan)

Description

In this study, computer simulations and experiments were used to verify the accuracy of a two-dimensional image registration program (program) for portal images that we previously developed. The program used a computed radiography cassette system and digitally reconstructed radiography images as planning images for external beam radiation therapy. Using this program, we also investigated the reason two-dimensional automatic image registration images experienced large misregistration in clinical practice using commercial image registration systems. Mutual information and normalized mutual information were used as the registration criteria. To investigate the influence of image background with or without a region of interest (ROI), results of image registrations were compared. Parameters of image registration were defined as translation in the horizontal and vertical directions (x and y, respectively) and rotation (θ) around the axis perpendicular to the x-y plane. There was no significant difference in image registration arising from the difference between mutual information and normalized mutual information. Image registration was improved with a ROI. Regardless of the registration criteria, errors in image registration with a ROI in the experimental study were <1.2 mm in directions x and y and <1.0 degree in rotation θ. We found that image registration required setting up as close to the planned position as possible. (author)

Availability note (English)

Available from http://dx.doi.org/10.6009/jjrt.2018_JSRT_74.8.779

Additional details

Additional titles

Original title (Japanese)
頭部ファントムを使用したCR ポータル画像とDRR 画像の位置照合プログラムの検証

Publishing Information

Journal Title
Nippon Hoshasen Gijutsu Gakkai Zasshi (Online)
Journal Volume
74
Journal Issue
8
Series
雑誌名:日本放射線技術学会雑誌
Journal Page Range
p. 779-788
ISSN
1881-4883

Optional Information

Notes
18 refs., 8 figs., 3 tabs.