Published November 1, 1975
| Version v1
Journal article
A 400 GeV shower counter
Description
Specifications and test data are given for a lead-scintillator electromagnetic shower counter used to detect beam photons at Fermi National Accelerator Laboratory. The response is independent of beam position to +- 1.4% rms over a 25 cmx25 cm area. Shower penetration losses reach 4.1% at 250 GeV but are easily calculable. The resolution is limited almost entirely by fluctuations in the number of shower tracks sampled: The full-width-at-half-maximum is approximately 43% Esup(1/2), where E is the incident energy in GeV. (Auth.)
Additional details
Identifiers
Publishing Information
- Journal Title
- Nuclear Instruments and Methods
- Journal Volume
- 129
- Journal Issue
- 1
- Series
- Nucl. Instrum. Methods.
- Journal Page Range
- 91-94
- ISSN
- 0029-554X
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 7237869
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ENERGY RESOLUTION; GEV RANGE 100-1000; PHOTOMULTIPLIERS; PHOTONS; PLASTIC SCINTILLATION DETECTOR; SHOWER COUNTERS; SHOWERS
- Descriptors DEC
- ELEMENTARY PARTICLES; ENERGY RANGE; GEV RANGE; MASSLESS PARTICLES; MEASURING INSTRUMENTS; PHOTOTUBES; RADIATION DETECTORS; RESOLUTION; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS