Published 1996 | Version v1
Journal article

A simple UV irradiator for low dose reassessment with LiF TLD-100

  • 1. Centro de Investigaciones Energeticas, Medioambientales y Tecnologicas (CIEMAT), Madrid (Spain)

Description

A simple and efficient UV irradiator specifically designed for dose reassessment by phototransferred thermoluminescence is described. The instrument permits the automatic irradiation of loose TLD chips for a preselected time and at a controlled temperature. The good phototransference characteristics allow the efficient reassessment of absorbed doses of the order of 0.2 mGy with good measurement performance, well within the limits required for personnel dosimetry. (Author)

Additional details

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
67
Journal Issue
4
Journal Page Range
p. 303-306.
ISSN
0144-8420
CODEN
RPDODE