Electron spectrometer designed for the internal conversion electron detection in nuclear reactions is described. The spectrometer efficiency is 10%, and energy resolution at the electron energy of 177 keV amounts to 2 keV. The range of measured electron energy can be varied in the process of measurement without vacuum breaking in the spectrometer. Calibration data and results of first experiments with the accelerator beam are presented