Published December 2006 | Version v1
Journal article

Precise positional measurement system in transcranial magnetic stimulation

  • 1. Japan Science and Technology Agency, Innovation plaza Hiroshima, Higashihiroshima, Hiroshima (Japan)
  • 2. Hiroshima City Univ., Graduate School of Information Sciences, Hiroshima, Hiroshima (Japan)
  • 3. Hiroshima City Univ., Faculty of Information Sciences, Hiroshima, Hiroshima (Japan)

Description

Transcranial magnetic stimulation (TMS) is a method for noninvasive stimulation of cerebral cortex, and it has contributed to clinical and basic researches of brain function. In order to estimate the accurate stimulating points of the cortex in TMS, precise measurement of the subject's head and the stimulating coil is necessary. In this study, we have developed the positioning TMS system with a three-dimensional (3-D) digitizer and a multi-articular system. We proposed a method for the accurate measurement of a subject's head and cortex, in which the location data of the subject's face surface captured by a 3-D digitizer were superimposed on the magnetic resonance imaging (MRI) data of the subject's face surface. Using this system, the precise estimation of the stimulated sites of the cortex in TMS was achieved. The validity of the system was verified by the experiment on the TMS of the motor cortex. (author)

Additional details

Publishing Information

Journal Title
Seitai Ikougaku
Journal Volume
44
Journal Issue
4
Journal Page Range
p. 674-679
ISSN