Published February 2012 | Version v1
Journal article

A beam monitor detector based on doped silica and optical fibres

  • 1. Albert Einstein Center for Fundamental Physics, Laboratory for High Energy Physics, University of Bern, Sidlerstrasse 5, CH-3012 Bern (Switzerland)
  • 2. Dipartimento di Scienze Fisiche, Università di Napoli Federico II, Complesso Universitario di Monte S. Angelo, I-80126 Napoli (Italy)
  • 3. Institute of Applied Physics, University of Bern, Sidlerstrasse 5, CH-3012 Bern (Switzerland)

Description

A beam monitor detector prototype based on doped silica fibres coupled to optical fibres has been designed, constructed and tested, mainly for accelerators used in medical applications. Scintillation light produced by Ce and Sb doped silica fibres moving across the beam has been measured, giving information on beam position, shape and intensity. Mostly based on commercial components, the detector is easy to install, to operate and no electronic components are located near the beam. Tests have been performed with a 2 MeV proton pulsed beam at an average current of 0.8 μA. The response characteristics of Sb doped silica fibres have been studied for the first time.

Availability note (English)

Available from http://dx.doi.org/10.1088/1748-0221/7/02/T02001

Additional details

Publishing Information

Journal Title
Journal of Instrumentation
Journal Volume
7
Journal Issue
02
Journal Page Range
p. T02001
ISSN
1748-0221