Published 1985 | Version v1
Journal article

Choice of scintillation detector sizes and correction energy in case of application of dual energy method to computerized tomography

  • 1. Tomskij Politekhnicheskij Inst. (USSR). Nauchno-Issledovatel'skij Inst. Ehlektronnoj Introskopii

Description

Technique of the choice of optimum energy for correcting tomographic projection energy inhomogeneity and corresponding optimum length of crystals when using the dual energy method in computerized tomography is presented. Results of numerical calculations are presented for concrete materials (aluminium, iron, water) along with the results of counting recording method

Additional details

Additional titles

Original title (Russian)
Выбор размеров сцинтиляционных детекторов и энергии коррекции при использовании метода дуальной энергии в вычислительной томографии

Publishing Information

Journal Title
Defektoskopiya
Journal Issue
no.9
Series
Defektoskopiya.
ISSN
0130-3082
CODEN
DEFKA

Optional Information

Notes
For English translation see the journal Soviet Journal of Nondestructive Testing (USA).