Published May 2008
| Version v1
Journal article
On-line analysis of hydrogen isotopes with thermal conductivity detector
Creators
- 1. China Academy of Engineering and Physics, Mianyang (China)
Description
On-line analysis of hydrogen isotopes with thermal conductivity detector(TCD) using H2 carrier gas was experimentally studied. Moreover, the influence of the sample gas pressure, deuterium concentration and HD content on measured results was investigated and discussed. The experimental results show that a good linear relation between TCD response to the partial pressure of deuterium is obtained. Various H2-D2 mixtures with 1%-90% deuterium concentration were used to verify this method. The error between measured value and theoretical value is in direct proportion to the HD content. The absolute error range is 0.000-0.025. (authors)
Additional details
Publishing Information
- Journal Title
- Journal of Nuclear and Radiochemistry
- Journal Volume
- 30
- Journal Issue
- 2
- Journal Page Range
- p. 98-102
- ISSN
- 0253-9950
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 41049934
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Descriptors DEI
- ACCURACY; DATA; DEUTERIUM; HYDROGEN; ISOTOPE RATIO; MIXTURES; PARTIAL PRESSURE; THERMAL CONDUCTIVITY
- Descriptors DEC
- DIMENSIONLESS NUMBERS; DISPERSIONS; ELEMENTS; HYDROGEN ISOTOPES; INFORMATION; ISOTOPES; LIGHT NUCLEI; NONMETALS; NUCLEI; ODD-ODD NUCLEI; PHYSICAL PROPERTIES; STABLE ISOTOPES; THERMODYNAMIC PROPERTIES
Optional Information
- Notes
- 4 figs., 4 tabs., 6 refs.