A comparison of patient doses in lumbar spine radiography from various X-ray units in Poland

  • 1. Instytut Medycyny Pracy w Przemysle Wlokienniczym i Chemicznym, Lodz (Poland)

Description

A number of diagnostic X-ray units in Poland were subjected to analysis. The air kerma on X-ray tables was measured for the X-ray parameters used routinely during lumbo-sacral spine radiography of standard adult patient. Measurements of air kerma were performed using TLD made of lithium-fluoride. The doses received by patients in different X-ray departments can differ more than two orders of magnitude (from 0.42 mGy up to 72.11 mGy), for the same X-ray examination. The dose can be significantly reduced if fluorescent screens made from components of rare earth are used. (author). 10 refs, 1 fig., 2 tabs