Published February 1989 | Version v1
Journal article

Design and imaging studies of a position sensitive photomultiplier based dynamic SPECT system

  • 1. Signal and Image Processing Institute, Dept. of Electrical Engineering - Systems, Univ. of Southern California, Los Angeles, CA (USA)
  • 2. Dept. of Radiology, Univ. of Southern California, Los Angeles, CA (USA)

Description

The design and image reconstruction aspects of a new generation of dynamic SPECT instrumentation utilizing multiple rings of position sensitive photomultiplier tubes (PSPMTs) are reported. A diverging cone-beam collimator is attached to each PSPMT to achieve a 22 cm diameter spherical field of view in a system designed for head imaging. The resolution with the diverging collimator is estimated to be approximately 20% worse than a parallel hole collimator. Iterative and direct cone-beam reconstruction algorithms have been investigated under different conditions of angular sampling and counting statistics. The ramifications of temporal resolution and signal to noise ratio are discussed in relation to dynamic SPECT imaging of labeled cisplatin for chemotherapy

Additional details

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
36
Journal Issue
1
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
1132-1137
ISSN
0018-9499
CODEN
IETNA

Conference

Title
IEEE nuclear science symposium.
Dates
9-11 Nov 1988.
Place
Orlando, FL (USA).

Optional Information

Secondary number(s)
CONF-881103--.