Published 1998 | Version v1
Report Restricted

A demonstration of the gross count tomographic gamma scanner (GC-TGS) method for the nondestructive assay of transuranic waste

Description

The authors examined the accuracy and sensitivity levels for three variations on the TGS method: the original TGS method using a high-purity germanium (HPGe) detector to measure net areas of full-energy gamma-ray peaks; a modified HPGe-detector method that uses net areas for the transmission analysis and the gross count TGS (GC-TGS) method for the emission analysis; and a NaI-detector method that uses the GC-TGS method exclusively. They found that while the accuracies of the methods were comparable, the GC-TGS method boosted the sensitivity per detector by a factor of approximately two for the HPGe GC variation and four for the NaI method. The implications for improved TGS scanner design are discussed

Availability note (English)

Available from INIS in electronic form; ALSO AVAILABLE FROM OSTI AS DE99002304; NTIS; INIS; US GOVT. PRINTING OFFICE DEP.

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Publishing Information

Imprint Pagination
13 p.
Report number
LA-UR--98-3248

Conference

Title
39. Institute of Nuclear Materials Management (INMM) annual meeting
Dates
26-30 Jul 1998
Place
Naples, FL (United States)

Optional Information