Published March 1, 1999 | Version v1
Journal article

A position-sensitive scintillation detector for two-dimensional angular correlation of annihilation radiation using metal-package position-sensitive photomultiplier tubes

  • 1. Institute of Physics, Graduate School of Arts and Sciences, University of Tokyo, 3-8-1 Komaba, Meguro-ku, Tokyo 153-8902 (Japan)
  • 2. Hamamatsu Photonics Co. Ltd., 314-5, Shimokanzo, Toyooka Village, Iwata-Gun, Shizuoka Pref. 438-0126 (Japan)

Description

We have constructed and tested a prototype of a new position sensitive γ-ray detector which consists of an array of 2.6x2.6x18 mm3 BGO scintillator blocks, a light guide, and four metal-package position-sensitive photomultiplier tubes (R5900-00-C8) recently developed by Hamamatsu Photonics Co. Ltd. Scalability of the detector of this type makes it possible to construct a larger detector using many PS-PMTs, which will be useful for the two-dimensional angular correlation of annihilation radiation apparatus

Additional details

Identifiers

PII
S0168900298012662;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
423
Journal Issue
2-3
Journal Page Range
p. 364-368
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 1999 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.