Description

It describes one method to provide high-precision clock and test-synchronization signal source for the 4πβ(LSC)-γ radioactivity digital coincidence platform data collection system. The circuit design based on PLL + DDS technology, PLL provides system clock with low jitter, while DDS generates sine wave for synchronization test. On PXI chassis, the ALTERA PCI-CORE is responsible for the communication between the plug-in boards with the zero slot. (authors)

Additional details

Publishing Information

Journal Title
Nuclear Electronics and Detection Technology
Journal Volume
34
Journal Issue
1
Journal Page Range
p. 40-44
ISSN
0258-0934

INIS

Country of Publication
China
Country of Input or Organization
China
INIS RN
49064817
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
ACCURACY; COMMUNICATIONS; DESIGN; GAMMA RADIATION; IONIZING RADIATIONS; RADIOACTIVITY; SIGNALS; SYNCHRONIZATION
Descriptors DEC
ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; RADIATIONS

Optional Information

Notes
9 figs., 2 tabs., 4 refs.