Modification of dose algorithm for Teledyne 9150 TLD system and performance test of the algorithm via intercomparison studies

  • 1. KAERI, Taejon (Korea, Democratic People's Republic of)

Description

To encourage dosimeter processors to evaluate personnel radiation dose more accurately, the Ministry of Science and Technology (MOST) of Korea issued an Ministerial Ordinance on the technical criteria on personnel dosimetry in August 1992 which was very similar to the ANSI N13.11 standard. Korea Atomic Energy Research Institute (KAERI) has modified the dose calculation algorithm of our Teledyne 9150 TLD system to meet the 8 test categories specified in the MOST Ordinance since 1991, and to provide a dosimeter performance test program to all domestic dosimeter processors. This paper describes the modified calculation algorithm, radiation fields used for modifying the existing algorithm, and presents intercomparison results for the performance of the modified algorithm with Atlan-Tech Inc., Oak Ridge National Lab. (ORNL), Japan Atomic Energy Research Institute (JAERI) and Australian Radiation Lab. (ARL). All intercomparison results showed a good agreement between irradiated and evaluated doses within the tolerance level of MOST Ordinance

Additional details

Publishing Information

Imprint Title
Fourth conference on radiation protection and dosimetry: Proceedings, program, and abstracts
Imprint Pagination
620 p.
Journal Page Range
p. 575-584.
Report number
ORNL/TM--12817

Conference

Title
43. Franz-Pacher colloquium on geomechanics.
Dates
13-14 Oct 1994.
Place
Salzburg (Austria).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
26046982
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S61: RADIATION PROTECTION AND DOSIMETRY; S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ALGORITHMS; CALCULATION METHODS; DOSE RATES; PERSONNEL DOSIMETRY; RADIATION DOSES; SAFETY STANDARDS; THERMOLUMINESCENT DOSIMETRY
Descriptors DEC
DOSIMETRY; STANDARDS

Optional Information

Secondary number(s)
CONF-9410171--.