Published May 1992 | Version v1
Journal article

Initial response of an automatic dew-point hygrometer using β-ray backscattering

  • 1. Saitama Univ., Urawa (Japan). Faculty of Engineering

Description

This paper describes the initial response of an automatic dew-point hygrometer using β-ray backscattering. The response time after starting of measurement was evaluated in which the dew point measured with the hygrometer reached a value of 90 percent of the correct dew point in its response. Theoretical calculation was made by means of the loop transfer function of a control system employed in the hygrometer. The response times ranged from 1.7 min to 16 min corresponding to the dew point of 47.5degC and -6degC, respectively, and agreed well with experimental ones. (author)

Additional details

Publishing Information

Journal Title
Radioisotopes (Tokyo)
Journal Volume
41
Journal Issue
5
Journal Page Range
p. 247-252.
ISSN
0033-8303
CODEN
RAISAB