Published May 1985 | Version v1
Journal article

Hight resolution Si(Li) X ray detector

  • 1. Beijing Nuclear Instrument Factory (China)

Description

This paper describes the fabrication technology of GL1221 type Si(Li) X ray detector core and the pulse light feedback colded preamplifier fitted on the detector. The energy resolution of the detector system is 165 eV (At 5.89 KeV Mn-K α X ray); the counting rate is 1020 cps, and the electronics noise is 104 eV. The performace of the detector keeps up with the business level of a foreign product of the same kind

Additional details

Publishing Information

Journal Title
Nucl. Electron. Detect. Technol.
Journal Volume
5
Journal Issue
3
Series
Nucl. Electron. Detect. Technol.
Journal Page Range
139-142
CODEN
HDYUE