Description

In precision radiotherapy, the intrafractional motion can cause a considerable uncertainty of the location of the tumor to be treated. An established approach is the expansion of the target volume to account for the motion. An alternative approach is couch-tracking, in which the patient is continually moved to compensate the intrafractional motion. However, couch-tracking itself might induce uncertainty of the patient's body position, because the body is non-rigid.

One hundred healthy volunteers were positioned supine on a robotic couch. Optical markers were placed on the torso of the volunteers as well as on the couch, and their positions were tracked with an optical surface measurement system. Using these markers, the uncertainty of the body position relative to the couch position was estimated while the couch was static or moving.

Over the included 83 healthy volunteers, the median of the uncertainty increased by 0.8 mm (SI), 0.4 mm (LR) and 0.4 mm (AP) when the couch moved.

Couch motion was found to increase the uncertainty of the body position relative to the couch. However, this uncertainty is one order of magnitude smaller than the intrafractional tumor motion amplitudes to be compensated. Therefore, even with body motion present, the couch-tracking approach is a viable option. The study was registered at ClinicalTrials.gov (NCT02820532) and the Swiss national clinical trials portal (SNCTP000001878). (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1361-6560/aaf361

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
64
Journal Issue
1
Journal Page Range
[9 p.]
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
52003642
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
CLINICAL TRIALS; NEOPLASMS; PATIENTS; RADIOTHERAPY
Descriptors DEC
DISEASES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; TESTING; THERAPY