Published 2010 | Version v1
Journal article

Chromatographic determination of radiochemical purity. Replacement of ITLC SG

  • 1. Universitaetsklinikum Carl Gustav Carus, Technische Univ. Dresden (Germany). Klinik und Poliklinik fuer Nuklearmedizin
  • 2. ROTOP Pharmaka AG, Radeberg (Germany)

Description

Thin layer chromatography is well established for quality control of radiopharmaceuticals. A convenient and widely used stationary phase are ITLC SG strips. However, the Pall Corporation stopped manufacturing of the silica gel impregnated glass fibre strips (ITLC SG). Material, Methods: As a replacement we tested silicic acid impregnated glass fibre strips from Varian (ITLC SA) and sufficient mobile phases. Results: The chromatography with these strips takes two to three times longer than with ITLC SG, but it is in an acceptable range. Only three mobile phases are necessary to test most of the common in-house made radiopharmaceuticals. Conclusion: The proposed method is suitable for routinely measuring the radiochemical purity of radiophamaceuticals.

Additional details

Additional titles

Original title (German)
Chromatographische Bestimmung der radiochemischen Reinheit von Radiopharmaka. Alternativen zu ITLC SG

Publishing Information

Journal Title
Nuklearmedizin
Journal Volume
49
Journal Issue
2
Journal Page Range
p. 73-77
ISSN
0029-5566
CODEN
NUKLDV