Analysis of lens absorbed dose in head CT scan using a male anthropomorphic phantom

  • 1. Federal University of Minas Gerais, Nuclear Engineering Department, Pampulha, 31270-901 Belo Horizonte, Minas Gerais (Brazil)
  • 2. Federal University of Minas Gerais, Anatomy and Imaging Department, 30130-100 Belo Horizonte, Minas Gerais (Brazil)

Description

This study aims to perform a dosimetric study on lens in head CT scan. A Ge CT scanner, model Discovery with 64 channels, was used for irradiated a male anthropomorphic phantom, model Alderson Rando and radiochromic film strips were used for dose measures. Two head scans were performed using the routine protocol used in the radiodiagnosis service, programming the appliance with a voltage of 120 kV, a current of 175 m A and a pitch of 0.984. Experiments were performed with the phantom in supine position and in supine with the head tilted. Radiochromic film strips were placed in the lenses and in nearby radiosensitive organs, such as: thyroid, pharynx, hypophysis, salivary glands, spinal cord and breasts for each scan. Recorded absorbed doses were among 1.80 and 19.23 mGy. Recorded absorbed doses were lower with the head tilted, mainly in the lenses. The analysis of noise indexes in the image of the head central slice presented acceptable values for soft tissues, less than 1%. (Author)

Files

51012325.pdf

Files (975.5 kB)

Name Size Download all
md5:cc0562821ff7077ab646af9cd9c41ab0
975.5 kB Preview Download

Additional details

Publishing Information

Publisher
Sociedad Mexicana de Irradiacion y Dosimetria
Imprint Place
Ciudad de Mexico (Mexico)
Imprint Pagination
10 p.
Report number
INIS-MX--3220

Conference

Title
18. international symposium on solid state dosimetry
Acronym
ISSSD 2018
Dates
24-28 Sep 2018
Place
Oaxaca, Oax. (Mexico)