Absolute efficiency calibration of HPGe detector by simulation method

  • 1. Radiation Safety Systems Division, Bhabha Atomic Research Centre, Mumbai (India)

Description

High resolution gamma ray spectrometry by HPGe detectors is a powerful radio analytical technique for estimation of activity of various radionuclides. In the present work absolute efficiency calibration of the HPGe detector was carried out using Monte Carlo simulation technique and results are compared with those obtained by experiment using standard radionuclides of 152Eu and 133Ba. The coincidence summing correction factors for the measurement of these nuclides were also calculated

Part of:
Proceedings of the thirty-third IARP international conference on developments towards improvement of radiological surveillance at nuclear facilities and environment: book of abstracts

Additional details

Publishing Information

Publisher
Indian Association for Radiation Protection
Imprint Place
Mumbai (India)
Imprint Title
Proceedings of the thirty-third IARP international conference on developments towards improvement of radiological surveillance at nuclear facilities and environment: book of abstracts
Imprint Pagination
314 p.
Journal Page Range
p. 179

Conference

Title
33. IARP international conference on developments towards improvement of radiological surveillance at nuclear facilities and environment
Acronym
IARPIC-2018
Dates
16-20 Jan 2018
Place
Mumbai (India)

INIS

Country of Publication
India
Country of Input or Organization
India
INIS RN
49034788
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CALIBRATION; GAMMA SPECTROSCOPY; HIGH-PURITY GE DETECTORS; PERFORMANCE TESTING; SIMULATION
Descriptors DEC
GE SEMICONDUCTOR DETECTORS; MEASURING INSTRUMENTS; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; SPECTROSCOPY; TESTING

Optional Information

Notes
1 ref., 1 fig.