Published August 1985 | Version v1
Report Restricted

Dosimetry of the portable blood irradiator

Description

A portable blood irradiator was developed at the Pacific Northwest Laboratory to evaluate the effects of chronic irradiation of the blood in suppressing graft rejection. The irradiator, designed to be worn on the arm or leg and be surgically connected in an arterio-venous shunt, uses beta radiation from activated thulium imbedded in a vitreous carbon matrix to reduce the number of lymphocytes circulating in the blood. The dose and energy spectra relative to the distance from and position around a prototype irradiator were measured by thermoluminescent dosimeters, ion chambers and photon spectroscopy. With computer simulations using those measurements, the shielding was redesigned to minimize the radiation dose to the patient and to the attending personnel and to minimize the weight of the irradiator. The new shielding design was incorporated into a new prototype, and the dose and spectral measurements were repeated, which confirmed the design improvements. 10 refs., 11 figs

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS, PC A03/MF A01 as DE85018216.

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Publishing Information

Imprint Pagination
28 p.
Report number
PNL-SA--13054

Conference

Title
27. annual meeting of the American Association of Physicists in Medicine.
Dates
11-15 Aug 1985.
Place
Seattle, WA (USA).

Optional Information

Secondary number(s)
CONF-8508125--1.