Published February 1993 | Version v1
Journal article

Perfusion of the human distal colon and rectum evaluated with endoscopic laser Doppler flowmetry

  • 1. Aker University Hospital, Oslo (Norway)

Description

The aim of this study was to evaluate methodologic aspects of colonoscopic laser Doppler flowmetry. A Periflux PF1d flowmeter, set to 4 kHz/0.2 sec, with an endoscopic probe was used. In 20 patients, with a median age of 70 years and without colonic disease, flux was recorded at 10, 40, 30, 20 and again at 10 cm from the anal verge. A median of three repeated recordings were made at each level to calculate average flux and spatial variation. Median flux was 158 perfusion units, and the coefficient of variation of repeated recordings 0.14. There was no regional variation, and no increase in flux at 10 cm from the start until the end of the procedure. Pressure of the probe against the bowel wall and severe distention significantly reduced the flux. The interference of light from the endoscopic light source on the flux could not be predicted. It differed with different light sources, and also with the length of probe coming out of the colonoscope - that is, the distance from the light to the measurement point. To avoid the problem, the light source should be turned off while recording. 19 refs., 4 figs

Additional details

Additional titles

Subtitle (English)
Methodologic aspects

Publishing Information

Journal Title
Scandinavian Journal of Gastroenterology
Journal Volume
28
Journal Issue
2
Journal Page Range
p. 104-108.
ISSN
0036-5521
CODEN
SJGRA4