Published January 1, 2002 | Version v1
Journal article

Characterisation of the IPNE Bonner sphere spectrometer by comparison with the PTB system

Description

An existing set of Bonner spheres of the Institute for Physics and Nuclear Engineering in Bucharest has been modified following the PTB design for application of an 3He-filled SP9 counter. By simple interpolations of the fluence responses of the PTB spheres, a preliminary fluence response matrix could be established for the IPNE Bonner sphere set. For further adjustments of the preliminary responses and for validation of the final fluence response matrix of the Romanian Bonner sphere spectrometer, calibration measurements with a reference 252Cf neutron source and joint measurements, including the PTB Bonner sphere spectrometer, were performed in a few neutron fields differing substantially in their spectral distributions. It is estimated that the integral neutron fluences and dose-equivalent values can now be determined with the Romanian spectrometer with uncertainties of about ±4% and ±8%, respectively

Additional details

Identifiers

PII
S016890020101381X;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
476
Journal Issue
1-2
Journal Page Range
p. 21-25
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2002 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.