Published 1999 | Version v1
Miscellaneous Open

Neutron dosimetry based on nuclear track etched detectors

  • 1. Grup de Fisica de les Radiacions. Departament de Fisica, Universitat Autonoma de Barcelona, E-08193 Bellaterra (Spain)

Description

In this work, the response of a neutron dosimeter based on plastic track detectors has been studied. The detector geometry used consists on a C R-39 detector 500 m thick plus either a Makrofol converter 300 μ m thick or air used as converter, for the study of the response to fast or thermal neutrons respectively. The possibility of using Makrofol as a high energy neutron dosemeter has also been studied. In order to validate the results obtained from Monte Carlo simulations, a set of irradiations to monoenergetic neutron beams has been performed at the Ptb and to realistic fields at Cadarache neutron irradiation facilities. An excellent agreement has been found between the simulated and the experimental values. The lower detection limit value found for C R-39 and fast neutrons was 60 μ Sv. (Author)

Availability note (English)

Available from INIS in electronic form

Files

31065605.pdf

Files (158.7 kB)

Name Size Download all
md5:e824a387d36ea5ca8a97bed11884313c
158.7 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
6 p.
Report number
INIS-MX--568

Conference

Title
2. International Conference; 12. National Congress on Solid State Dosimetry
Dates
22-24 Sep 1999
Place
Mexico City (Mexico)