Published May 2014 | Version v1
Journal article

Standardization of 237Np

Description

The standardization of 237Np was investigated. The certified massic activity for 237Np was obtained by 4παβ liquid scintillation (LS) counting with correction for the 233Pa daughter using the CIEMAT/NIST efficiency tracing method using a 3H standard. Confirmatory measurements were also performed by high-resolution HPGe gamma-ray spectrometry, and by 4παβ(LS)-γ(NaI) anticoincidence counting. All results agree within the respective method's uncertainties. It was confirmed that the 237Np/233Pa radioactive equilibrium is disturbed when making dilutions and/or removing aliquots. - Highlights: • Standardization of 237Np by 4παβ liquid scintillation counting and confirmatory methods. • All results agree within the respective method's uncertainties. • 237Np/233Pa radioactive equilibrium disturbed when making dilutions and/or removing aliquots

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2013.11.074

Additional details

Identifiers

DOI
10.1016/j.apradiso.2013.11.074;
PII
S0969-8043(13)00523-X;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
87
Journal Page Range
p. 269-273
ISSN
0969-8043
CODEN
ARISEF

Conference

Title
19. international conference on radionuclide metrology and its applications
Dates
17-21 Jun 2013
Place
Antwerp (Belgium)

Optional Information

Copyright
Copyright (c) 2013 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.