Published 1981 | Version v1
Journal article

CAMAC module for waterflow calorimetry

  • 1. Lawrence Berkeley Lab, CA, USA

Description

Waterflow calorimetry would seem to be a natural diagnostic for Neutral Beam Injectors: it requires simple transducers, is non-destructive and directly measures the 'bottom line' number of delivered neutral power. Real Time Systems Group, Lawrence Berkeley Laboratory has developed a micro-processor based CAMAC module for doing waterflow calorimetry which gives an energy measure accurate to better than 2% within 1 minute after a shot. The modules are relatively inexpensive and have been proven in normal operation at LBL's Neutral Beam System Test Facility. This paper describes the module and experience gained in its development and operation

Additional details

Publishing Information

Journal Title
Proc. Symp. Eng. Probl. Fusion Res.
Journal Issue
p. 1078-1081
Series
Proc. Symp. Eng. Probl. Fusion Res.
ISSN
0145-5958

Conference

Title
9. symposium on engineering problems of fusion research.
Dates
26 - 29 Oct 1981.
Place
Chicago, IL (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
14766202
Subject category
S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ALGORITHMS; CALORIMETRY; CAMAC SYSTEM; DIAGNOSTIC TECHNIQUES; NEUTRAL ATOM BEAM INJECTION; NEUTRAL BEAM SOURCES; NUCLEAR INSTRUMENT MODULES
Descriptors DEC
BEAM INJECTION

Optional Information

Secondary number(s)
CONF-811040--.