Published July 2024 | Version v1
Journal article

A segmentation-based automated calculation of patient size and size-specific dose estimates in pediatric computed tomography scans

  • 1. Department of Biomedical Imaging, Advanced Medical and Dental Institute, Universiti Sains Malaysia(Malaysia)
  • 2. Department of Computer Science, School of Computer Science, Universiti Sains Malaysia, Penang (Malaysia)

Description

Size-specific dose estimates (SSDE) have been introduced into computed tomography (CT) dosimetry to tailor patients' unique sizes to facilitate accurate CT radiation dose quantification and optimization. The purpose of this study was to develop and validate an automated algorithm for the determination of patient size (effective diameter) and SSDE. A MATLAB platform was used to develop software of algorithms based on image segmentation techniques to automate the calculation of patient size and SSDE. The algorithm was used to automatically estimate the individual size and SSDE of four CT dose index phantoms and 80 CT images of pediatric patients comprising head, thorax, and abdomen scans. For validation, the American Association of Physicists in Medicine (AAPM) manual methods were used to determine the patient's size and SSDE for the same subjects. The accuracy of the proposed algorithm in size and SSDE calculation was evaluated for agreement with the AAPM's estimations (manual) using Bland-Altman's agreement and Pearson's correlation coefficient. The normalized error, system bias, and limits of agreement (LOA) between methods were derived. The results demonstrated good agreement and accuracy between the automated and AAPM's patient size estimations with an error rate of 1.9% and 0.27% on the patient and phantoms study, respectively. A 1% percentage difference was found between the automated and manual (AAPM) SSDE estimates. A strong degree of correlation was seen with a narrow LOA between methods for clinical study (r > 0.9771) and phantom study (r > 0.9999). The proposed automated algorithm provides an accurate estimation of patient size and SSDE with negligible error after validation. (author)

Additional details

Identifiers

Publishing Information

Journal Title
Journal of Medical Physics
Journal Volume
49
Journal Issue
3
Journal Page Range
p. 456-463
CODEN
JMPHFE

INIS

Country of Publication
India
Country of Input or Organization
India
INIS RN
55092837
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; CT-GUIDED RADIOTHERAPY; DOSE LIMITS; DOSE RATES; DOSIMETRY; PHANTOMS
Descriptors DEC
MEDICINE; MOCKUP; NUCLEAR MEDICINE; RADIOLOGY; RADIOTHERAPY; SAFETY STANDARDS; STANDARDS; STRUCTURAL MODELS; THERAPY