Published November 2019 | Version v1
Journal article

Interferences in the determination of lutetium using neutron activation analysis

  • 1. University of Texas, Austin, TX (United States). Nuclear Engineering Teaching Lab

Description

We have investigated the severity of interferences in the determination of lutetium in neutron activation analysis. The preferred reaction makes use of the large radiative capture cross section of 2090 b for 176Lu(n,γ)177Lu, which emits a 208.4 keV gamma emission (t1/2 = 6.73 days). Activation of 176Lu simultaneously produces 177mLu, which also emits at 208.4 keV (t1/2 = 160.4 days). We found that these convolutions did not produce significant error in lutetium analysis for a sample of NIST 1633c, even for longer decay times used to reduce the 239Np interference at 209.8 keV. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
322
Journal Issue
2
Journal Page Range
p. 1201-1205
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
12 refs.