Published June 5, 2012 | Version v1
Journal article

Neutron beam imaging at neutron spectrometers at Dhruva

  • 1. Solid State Physics Division, Bhabha Atomic Research Centre, Trombay, Mumbai 400 085 (India)

Description

A low efficiency, 2-Dimensional Position Sensitive Neutron Detector based on delay line position encoding is developed. It is designed to handle beam flux of 106-107 n/cm2/s and for monitoring intensity profiles of neutron beams. The present detector can be mounted in transmission mode, as the hardware allows maximum neutron transmission in sensitive region. Position resolution of 1.2 mm in X and Y directions, is obtained. Online monitoring of beam images and intensity profile of various neutron scattering spectrometers at Dhruva are presented. It shows better dynamic range of intensity over commercial neutron camera and is also time effective over the traditionally used photographic method.

Additional details

Identifiers

Publishing Information

Journal Title
AIP Conference Proceedings
Journal Volume
1447
Journal Issue
1
Journal Page Range
p. 491-492
ISSN
0094-243X
CODEN
APCPCS

Conference

Title
56. DAE solid state physics symposium 2011
Dates
19-23 Dec 2011
Place
Kattankulathur, Tamilnadu (India)

Optional Information

Notes
(c) 2012 American Institute of Physics