Published October 2012 | Version v1
Journal article

Development of a signal processing system to control a tumor positioning system based on annihilation gamma-rays measured using several pairs of scintillation detectors

  • 1. Toyama National College of Technology, Toyama (Japan)
  • 2. Hokkaido University, Sapporo, Hokkaido (Japan)

Description

The authors have been studying a new tumor tracking system for stereotactic radio surgery, composed of several pairs of scintillation detectors that detect annihilation gamma rays from radioactive medicine, such as 18F-FDG, in tumor tissue. For that purpose, a signal processing system was developed based on an ASIC (MPX-08, Nova), which can output the logical pulses when the signals from the photomultiplier tubes (PMTs) are above the preset discrimination level. Results obtained experimentally using a 22Na point source show that the system can control erroneous irradiation outside the expected region to less than 1.8%, even when the source moved with speed of 15 mm/s. Moreover, the computer simulation showed that the system can work with erroneous irradiation of less than several percent, even for volumetric sources, considering the three-dimensional distribution of 18F-FDG near a lung tumor. (author)

Availability note (English)

Available from http://dx.doi.org/10.15669/pnst.3.104

Additional details

Identifiers

Publishing Information

Journal Title
Progress in Nuclear Science and Technology
Journal Volume
3
Journal Page Range
p. 104-108
ISSN
2185-4823

Conference

Title
6. international symposium on radiation safety and detection technology
Acronym
ISORD-6
Dates
12-14 Jul 2011
Place
Langkawi (Malaysia)

Optional Information

Notes
8 refs., 10 figs., 1 tab.