Left ventricular global myocardial strain assessment comparing the reproducibility of four commercially available CMR-feature tracking algorithms
Creators
- 1. Universidad de Salamanca, y CIBERCV, Servicio de Cardiologia, Hospital Universitario de Salamanca, Instituto de Investigacion Biomedica de Salamanca (IBSAL), Facultad de Medicina, Salamanca (Spain)
- 2. KU Leuven - University of Leuven, Department of Imaging and Pathology, Leuven (Belgium)
- 3. KU Leuven - University of Leuven, Laboratory on Cardiovascular Imaging and Dynamics, Department of Cardiovascular Sciences, Leuven (Belgium)
- 4. KU Leuven - University of Leuven, Department of Cardiovascular Sciences, Leuven (Belgium)
Description
To compare the reproducibility of cardiovascular magnetic resonance feature-tracking (CMR-FT) packages to assess global left ventricular (LV) myocardial strain. In 45 subjects (i.e. 15 controls, 15 acute myocardial infarction, 15 dilated cardiomyopathy patients), we determined inter-vendor, inter-observer (two readers) and intra-observer reproducibility of peak systolic global radial, circumferential and longitudinal strain (GRS, GCS and GLS, respectively) comparing four commercially available software packages. Differences between vendors were assessed with analysis of variance (ANOVA), between observers and readings with intraclass correlation coefficient (ICC) and coefficient of variation (CV). The normalised end-diastolic volume was 91, 77 and 119 ml/m2 (median, Q1, Q3) and ejection fraction was 41 ± 14%, range 12-67%. Global longitudinal strain (GLS), global circumferential strain (GCS) and global radial strain (GRS) values were 13.9% ± 5.4% (3.9-23.8%), 12.2% ± 5.8% (1.0-25.1%) and 32.0% ± 14.7 (3.6-67.8%), respectively. ANOVA showed significant differences between vendors for GRS (p < 0.001) and GLS (p = 0.018), not for GCS (p = 0.379). No significant bias was found for both intra- and inter-observer variability. The ICC for inter- and intra-observer reproducibility ranged 0.828-0.991 and 0.902-0.997, respectively. The CV, however, ranged considerably, i.e. 4.0-28.8% and 2.8- 27.7% for inter- and intra-observer reproducibility, respectively. In particular, for GRS differences in CV values between vendors were large, i.e. 5.2-28.8% and 2.8-27.7%, for inter- and intra-observer reproducibility, respectively. In a cohort of subjects with a wide range of cardiac performances, GRS and GLS values are not interchangeable between vendors. Moreover, although intra- and inter-observer reproducibility amongst vendors is excellent, some vendors encounter problems to reproducibly measure global radial strain. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00330-018-5538-4Additional details
Identifiers
Publishing Information
- Journal Title
- European Radiology
- Journal Volume
- 28
- Journal Issue
- 12
- Journal Page Range
- p. 5137-5147
- ISSN
- 0938-7994
- CODEN
- EURAE3
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 50010129
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ALGORITHMS; CARDIOVASCULAR SYSTEM; COMPARATIVE EVALUATIONS; COMPUTER CALCULATIONS; COMPUTER CODES; IMAGE PROCESSING; MYOCARDIAL INFARCTION; NMR IMAGING; STRAINS; TIME DEPENDENCE
- Descriptors DEC
- CARDIOVASCULAR DISEASES; DIAGNOSTIC TECHNIQUES; DISEASES; EVALUATION; MATHEMATICAL LOGIC; PROCESSING