Reproducibility of volume and densitometric measures of emphysema on repeat computed tomography with an interval of 1 week
Creators
- 1. David Geffen School of Medicine, UCLA, Center for Computer Vision and Imaging Biomarkers, Los Angeles, CA (United States)
- 2. David Geffen School of Medicine, UCLA, Center for Computer Vision and Imaging Biomarkers, Department of Radiological Sciences, Los Angeles, CA (United States)
Description
The reproducibilities of CT lung volume and densitometric measures of emphysema were assessed over 1 week. The influence of breathhold on reproducibility was assessed. HRCT was performed on 44 subjects at inspiration on two visits with a 7-day interval. CT lung volume, relative area below -950HU (RA950-raw), and 15th percentile density (PD15-raw) were computed. Volume correction was used to obtain RA950-adj and PD15-adj. Reproducibilities between visits were assessed using concordance correlation coefficient (CCC) and repeatability coefficient (RC). Reproducibilities were compared between raw and adjusted measures. Differences between visits were computed for volume and density measures. Correlations were computed for density differences versus volume difference. Subgroup analysis was performed using a 0.25 L volume difference threshold. High CCC were observed for all measures in full group (CCC > 0.97). Reproducibilities of volume (RC = 0.67 L), RA950-raw (RC = 2.3%), and PD15-raw (RC = 10.6HU) were observed. Volume correction significantly improved PD15 (RC = 3.6HU) but not RA950 (RC = 1.7%). RA950-raw and PD15-raw had significantly better RC in <0.25 L subgroup than ≥0.25 L. Significant correlations with volume were observed for RA950-raw and PD15-raw (R 2 > 0.71), but not RA950-adj or PD15-adj (R 2 < 0.11). Good breathhold and RA950 reproducibilities were achieved. PD15 was less reproducible but improved with volume correction or superior breathhold reproduction. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00330-011-2277-1Additional details
Identifiers
Publishing Information
- Journal Title
- European Radiology
- Journal Volume
- 22
- Journal Issue
- 2
- Journal Page Range
- p. 287-294
- ISSN
- 0938-7994
- CODEN
- EURAE3
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 43028746
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BREATH; COMPUTERIZED TOMOGRAPHY; DENSITY; EMPHYSEMA; IMAGE PROCESSING; LUNGS; REGRESSION ANALYSIS; RESPIRATION; VOLUME
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; DISEASES; MATHEMATICS; ORGANS; PATHOLOGICAL CHANGES; PHYSICAL PROPERTIES; PROCESSING; RESPIRATORY SYSTEM; RESPIRATORY SYSTEM DISEASES; STATISTICS; TOMOGRAPHY