Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the NRC, Canada and the BIPM in mammography x-rays

  • 1. BIPM Bureau International des Poids et Mesures, Pavillon de Breteuil, F-92312 Sevres cedex (France)
  • 2. NRC National Research Council Canada 1200 Montreal Road, Ottawa, Ontario K1A 0R6 (Canada)

Description

A key comparison has been made between the air-kerma standards of the NRC and the BIPM in mammography x ray beams. The results show the standards to be in agreement at the level of the combined standard uncertainty of 3.2 parts in 103. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database. (authors)

Availability note (English)

Available from doi: https://doi.org/10.1088/0026-1394/57/1A/06011

Additional details

Publishing Information

Journal Title
Metrologia
Journal Volume
57
Journal Issue
no.1A
Journal Page Range
p. 06011.1-06011.13
ISSN
0026-1394

Optional Information

Notes
10 refs.